By Benjamin Dannan and Steven Sandler, Picotest.com
High-speed Printed Circuit Board (PCB) design requires well-designed Power Delivery Networks (PDN) to support today's FPGAs and custom mixed-signal ASICs. The PDN contains important impedance information that can tell a designer how a system will react to dynamic currents and the impact of PCB layout. Considering the PDN is the transmission line between the Voltage Regulator Module (VRM) and the load, it is then important to make the starting point for a good PDN design at the VRM.
Today, VRMs need to supply power to multiple VDD cores to support FPGAs and/or custom ASICs using multi-gigabit ethernet, PCIe, and DDR memory interfaces. With that being said, vendor information for a VRM’s output impedance is not available and not always accurate when it is supplied. Further, measuring ultra-low impedance on multiple VRMs or multi-topology DC-DC regulators is a challenge for any design engineer.
The 2-port shunt-through impedance measurement is the gold standard for measuring a VRM’s output impedance in the microOhm and milliOhm region [1]. However, it is not always possible to make these measurements without direct SMA connections to the PCB or Device Under Test (DUT). Therefore, when a designer makes these types of measurements with a Vector Network Analyzer (VNA), the method of connecting the DUT requires attention to detail to ensure inductance and various error sources are minimized to allow an accurate measurement. To get the most out of your VNA, you need to use the right probes and accessories to ensure your measurement is successful. With a browser probe like the P2102A, you can quickly characterize multiple VRMs to ensure stability or even check if your model is accurate during your initial PDN design.
The Picotest P2102A 2-port PDN transmission line probe is a browser probe that achieves a very low inductance at the tip to mitigate space constraints on a dense PCB while eliminating the need to solder coax, add additional SMA connections or other test points necessary for impedance measurements. It is especially useful when there are dozens of rails to assess, and there isn’t time or PCB iterations available to provide test point implementations for each. Repeated measurements are simplified because the connection is by simply touching the tip to an output capacitor. This browser probe comes with 4 probe tips to allow measurement across a variety of SMD packages on a PCB, such as 1206, 0805, 0603, or 0402. The P2102A probe tips are available with 1X, 2X, 5X, and 10X attenuations as well as a DC Block option. This allows flexibility for users to measure across a wide range of voltages. For instance, the 2X probe can measure 6Vrms without DC blocks. The tradeoff is the attenuation increases the impedance floor. In short, this 2-port P2102A probe is best suited for VRM, power plane, and decoupling measurements. An added benefit is that you can use Non-Invasive Stability Measurement (NISM) to assess power supply’s stability at the same time as you assess the PDN [2].
The goal of this article is to show design and test engineers the process of how to set up and use the Keysight E5061B VNA with the Picotest P2102A browser probe to accurately measure the impedance of any VRM or power rail efficiently and quickly. This will also show you how to use this browser probe as a quick GO/NO-GO tester. In this application note, two DUTs will be measured as detailed by the process shown in Figure 1.
Test Equipment List
Description | Model | QTY |
Vector Network Analyzer | Keysight E5061B | 1 |
Picotest P2102A-1X | 1 | |
Picotest J2102B-N | 1 | |
2-Port Probe Adapter Panel | Picotest J2160A | 1 |
BNCJ/BNCJ-250 | 1 | |
N Male to BNC Female Adapter | Pasternack PE9002 | 2 |
Picotest 3DPP-200 | 1 | |
VRM Demo Boards (Flat and Varying Impedance) | LM20143B | 1 |
VRM - Infineon PS5401 Eval (DUT) | EVAL_PS5401-INT | 1 |
Calibration Board/Substrate | Included in Picotest P2102A kit | 1 |
*Note: The Picotest LM20143 test board may be referred to as Flat DUT and Not Flat (NF) DUT throughout this document.
Measurement Setup
The P2102A-2X probe tips include 50 Ω series resistors (Rs), which can be set up/accounted for in the E5061B Calibration menu. Figures 5 and 6 provide a depiction of how the DUT is connected to the 2-port P2102A probe with the E5061B S-parameter ports.
For the other P2102A-#X probe tip models, set Rs as defined below in the E5061B VNAs:
P2102A-1X - Rs = 0Ω
P2102A-5X - Rs = 200Ω
P2102A-10X - Rs = 450Ω
With the Keysight E5061B VNA, there are two possible 2-Port Shunt-Through measurement setups with the P2102A.
One setup includes the Picotest J2102B common mode transformer, connected to ports 1 and 2 on the E5061B S-parameter ports, as shown in Figure 5. The necessary test equipment for this setup is shown in Figure 2.
The shunt-thru configuration inherently suffers a ground loop error at low frequencies. The current flowing through the cable shield of the connection to channel 2 ground introduces a measurement error that can become significant at frequencies between 10 kHz to 100 MHz when measuring very low impedance values. To reduce the ground loop error at low frequencies, use a ground isolator or common mode transformer (e.g., J2102B) or an active isolation device such as the J2113A [12].
The other measurement setup includes the Picotest J2160A 2-port probe adapter, connected to the T/R ports low-frequency VNA BNC ports on the E5061B, as shown in Figure 6. The necessary test equipment for this setup is shown in Figure 3.
The Picotest J2160A Probe Adapter provides a low-noise, compact solution when using the E5061B T/R ports in a 2‐port shunt-through measurement. It converts the instruments’s three GP ports to the 2 ports necessary to perform a 2-port shunt-through measurement. The T/R ports are desirable for low-frequency 2‐port measurements, since the receiver port is semi‐floating, allowing low impedance measurements without the use of a coaxial common mode transformer such as the J2102B. The floating ports allow milliohm measurements even at very low frequencies and up to the 30 MHz range of the T/R ports.
2-port probe users generally own a probe station; however, the commitment to setting up the camera, microscope, calibration, etc., is a much bigger commitment than sometimes necessary for simple VRM impedance or even stability measurements. This is where the P2102A browser probe provides a great option on-the-go, as depicted by Figures 7 and 8.
Prior to making any measurements, you should ensure calibration of the setup is performed by using the proper calibration method. The E5061B VNA has multiple calibration options shown in Table 4-1 of the E5061B user manual. Proper calibration is critical since it corrects for contact resistance, tip inductance, coupling, and thermocouple effects. For the highest accuracy measurements, Picotest’s recommendation is to use the Full 2-Port Calibration which includes SHORT-OPEN-LOAD-THRU-ISOLATION calibration method with the E5061B. This ensures high accuracy with a 12-term error correction model by requiring 8 standard measurements shown by Figure 18. An example of why this is important is shown later in this document. In addition, to ensure consistent contact resistance, it is recommended that a probe holder be used during calibration and measurement. There are other calibration options available on the E5061B such as Enhanced Response Calibration which includes an OPEN-SHORT-LOAD-THRU calibration method to provide a 5-term error correction model, as well as a Response Calibration to provide a 2-term error correction model which includes either an OPEN-SHORT-LOAD or THRU-LOAD calibration.
Figures 9 and 10 provide a depiction of the probe tip location on each DUT. For Figure 9, an 0603 capacitor (C42) was removed prior to measurement and the 0603 probe tip is then used as shown; however, it is not necessary to remove the capacitor to make this measurement. For identification purposes, the probe head side with a label indicates the positive signal side.
Measurement Results
As shown by the results in Figures 13 and 16, it is possible to accurately measure below 1 mΩ with the P2102A browser probe, using either measurement setup.
P2102A Calibration and Setup Checklist with E5061B
Calibration of the testing setup for the shunt-through impedance with series resistance measurement with both the J2102B common mode transformer and the J2160A 2-port probe adapter.
2-Port Calibration Method with the E5061B and J2102B Common Mode Transformer
After powering on the E5061B, with cables connected for calibration as shown in Figure 7, follow the steps below to calibrate your measurement setup prior to making measurements on your DUT.
Step 1: (On the E5061B) Press > Cal button
Step 2: Set > Rs, by specifying under Z0 (as shown by Figure 17)
*Note: Rs only needs to be specified for P2102A-2X, 5X, 10X probes. For 1X probes this step can be ignored. For DC Block use 1X settings.
Step 3: Select > Calibrate
Step 4: Select > 2-Port Cal*
Step 5: Select > Reflection
Step 6: Place Probe on calibration substrate as appropriate
Step 7: Select > Port1 Open
✔ will appear next to the ‘Port1 Open’ LSK
Step 8: Select > Port2 Open
✔ will appear next to the ‘Port2 Open’ LSK
Step 9: Repeat for all 6 Reflection calibration steps until ✔ are reflected, as shown in Figure 19. Reference Figure 18 to complete all Full 2-port calibration steps.
Step 10: Select > Return
Step 11: Select > Transmission
Step 12: Select > Port 1-2 Thru
✔ will appear next to the ‘Port 1-2 Thru’ LSK
Step 13: Select > Return
Step 14: Select > Isolation (Optional)
Step 15: Select > Port 1-2 Isol
✔ will appear next to the ‘Port 1-2 Isol’ LSK
Step 16: Select > Return
*Note: Full 2-port calibration is completed when ✔ is shown for Reflection, Transmission, Isolation, as depicted by figure 20.
Step 17: Select > Done
Step 18: Select > Return
2-Port Calibration Method with the E5061B and J2160A 2-Port Probe Adapter
After powering on the E5061B, with cables connected for calibration as shown in Figure 8, follow the steps below to calibrate your measurement setup prior to making measurements on your DUT.
Step 1: (On the E5061B) Press > Meas button
Step 2: Select > Impedance Analysis Menu
Step 3: Select > Method Port 1 Refl
Step 4: Select > GP Shunt T 50Ω, R 50Ω
Step 5: Select > |Z|
Step 6: Select > Gain-Phase Setup
Step 7: Set > T input Z = 50 Ω
Step 8: Set > R Input Z = 50 Ω
Set the attenuation for T Attenuator & R Attenuator as needed
Step 9: Press > Cal button
Step 10: Set > Rs, by specifying under Z0 (as shown by Figure 17)
*Note: Rs only needs to be specified for P2102A-2X, 5X, 10X probes. For 1X probes this step can be ignored.
Step 11: Select > Calibrate
Step 12: Select > Response (Open)
Step 13: Ensure Select Port > GP Port*
Step 14: Place Probe on calibration substrate as appropriate
Step 15: Select > Open(f)
Step 16: Place Probe on calibration substrate as appropriate
Step 17: Select > Load (Optional)
Step 18: Select > Done
Step 19: Select > Return
Step 20: Repeat for Response (Short) and Response (Thru) calibration steps until ✔ are reflected. Reference Figure 18 to complete all Full 2-port calibration steps.
Checklist for Shunt-Through Measurement of the DUT with J2102B Common Mode Transformer
With the DUT connected to the E5061B to make a shunt-through measurement as shown in Figure 7.
Step 1: (On the E5061B) Press > Meas button
Step 2: Select > Impedance Analysis Menu
Step 3: Select > Method Port 1 Refl
Step 4: Select > Port 1-2 Shunt
Step 5: Select > |Z|
Step 6: Press > Scale button
Set scale or select ‘Auto Scale’ as appropriate
Step 7: Select > Y-axis Log
Step 8: Select > Log
Step 9: Press > Display button
Set title labels and display colors as appropriate
Step 10: Press > Ave button
Step 11: Set > IF Bandwidth
*Note: The lower the bandwidth ⇒ higher resolution, but also the slower the measurement. Typically, a good starting point is 1 kHz.
Step 12: Press > Start button
Step 13: Set Start frequency for measurement
Step 14: Press > Stop button
Step 15: Set Stop frequency for measurement
Step 16: Press > Sweep Setup button
Step 17: Select > Power
Step 18: Set Power dBm level
Step 19: Select > Return
Step 20: Select > Sweep Type
Step 21: Select > Log Freq
Step 22: Select > Return
Step 23: Select > Return (to return to E5061B Menu)
Begin capturing measurements of the DUT…
Checklist for Shunt-Through Measurement of the DUT with J2160A
With the DUT connected to the E5061B to make a shunt-through measurement as shown in Figure 8…
Step 1: (On the E5061B) Press > Meas button
Step 2: Select > Impedance Analysis Menu
Step 3: Select > Method Port 1 Refl
Step 4: Select > GP Shunt T 50Ω, R 50Ω
Step 5: Select > |Z|
Step 6: Select > Gain-Phase Setup
Step 7: Set > T input Z = 50 Ω
Step 8: Set > R Input Z = 50 Ω
*Note: If the signal being measured is noisy, decrease the attenuation (T Attenuator & R Attenuator) from 20dB as needed.
Step 9: Select > Return
Step 10: Press > Scale button
Set scale or select ‘Auto Scale’ as appropriate
Step 11: Select > Y-axis Log
Step 12: Select > Log
Step 13: Press > Display button
Set title labels and display colors as appropriate
Step 14: Press > Ave button
Step 15: Set > IF Bandwidth
*Note: The lower the bandwidth ⇒ higher resolution, but also the slower the measurement. Typically, a good starting point is 1kHz.
Step 16: Press > Start button
Step 17: Set Start frequency for measurement
Step 18: Press > Stop button
Step 19: Set Stop frequency for measurement
Step 20: Press > Sweep Setup button
Step 21: Select > Power
Step 22: Set > Power dBm level
*Note: In Gain-Phase Setup Menu, adjust attenuation levels as needed to prevent overload. This can be done in combination with the Power level.
Step 23: Select > Return
Step 24: Select > Sweep Type
Step 25: Select > Log Freq
Step 26: Select > Return
Step 27: Select > Return (to return to E5061B Menu)
Begin capturing measurements of the DUT…
Measurement accuracy can be further improved by using the full 2-port calibration method versus only the Thru calibration method. The Thru calibration method does not account for inductive coupling errors. Figures 21 and 22 provide two examples comparing the measurement error due to inductive coupling when using the full 2-port calibration method versus Thru method.
As shown by Figure 21, a 37 mΩ decrease is observed at 10 MHz with the full 2-port calibrated measurement versus with the Thru calibrated measurement on the LM20143 Not Flat DUT. A similar result is also seen by Figure 22, where a 21 mΩ decease at 5 MHz is observed by using the full 2-port calibration method in comparison to the Thru calibration method.
Conclusion
The 2-port shunt-through impedance testing method shown here is the gold standard for measuring a VRM's output impedance and the control loop gain (phase) stability performance.
The P2102A browser probe allows you to very simply, quickly, and accurately make 2-port impedance measurements that help better design your PDN and provide quick GO/NO-GO testing. When impedances are measured, not only can we determine a VRM’s stability, but we can also determine what the power supply distribution network is composed of. We can even create a highly accurate model of the VRM from the impedance measurement that includes time domain, frequency domain and even EMI-related data. We can tell which parts of the impedance are based on control loop performance and which parts are based on printed circuit board and/or decoupling performance. With good calibration, as demonstrated by the results in Figure 22, it is possible to achieve an impedance measurement below 1mΩ with the P2102A probe and the E5061B VNA.
If you really want to measure planes or high-frequency impedance, the P2104A 1-Port Probes [13] can be an even better option.
As shown in this document, to make low-impedance measurements with an E5061B VNA with the P2102A browser probe:
It is essential to have the correct (low inductance) probes, high-quality cables, and the common mode transformer (J2102B).
You must properly calibrate your setup, as this is essential to achieve high fidelity measurement and eliminate sources of test setup errors.
A consistent, repeatable contact resistance that applies consistent tip pressure is necessary. This can be done with a probe holder like the N2787A shown, ClampMan [14], or one of the probe holders from PacketMicro.